跳到主要內容

臺灣博碩士論文加值系統

(216.73.216.60) 您好!臺灣時間:2026/08/05 01:13
字體大小: 字級放大   字級縮小   預設字形  
回查詢結果 :::

詳目顯示

我願授權國圖
: 
twitterline
研究生:陳瑾瑩
研究生(外文):Chen, Chin-Yin
論文名稱:8-Hydroxyquinoline(8-HQ)固相萃取濃縮法及原子吸收光譜儀分析河口水重金屬之研究
論文名稱(外文):Determination of Trace Metals in Estuarine Water by 8-Hydroxyquinoline Solid-Phase Extraction and Graphite Atomic Absorption Spectrometry
指導教授:陳永祺
指導教授(外文):Chen Yung Chi
學位類別:碩士
校院名稱:中國文化大學
系所名稱:應用化學研究所
學門:自然科學學門
學類:化學學類
論文種類:學術論文
論文出版年:1998
畢業學年度:86
語文別:中文
論文頁數:2
中文關鍵詞:石墨爐式原子吸收光譜儀修飾劑
外文關鍵詞:8-HydroxyquinolineGraphite Atomic Absorption Spectrometrymodifier
相關次數:
  • 被引用被引用:0
  • 點閱點閱:189
  • 評分評分:
  • 下載下載:0
  • 收藏至我的研究室書目清單書目收藏:0
摘 要

本研究是利用8-Hydroxyquinoline (8-HQ)來萃取濃縮河口水中的重金屬
元素Cu、Cd,並以Hitachi Z-8270石墨爐式原子吸收光譜(GAAS,
Graphite Atomic Absorption spectrometry)來偵測濃縮後的水樣。首先
將合成後的8-HQ特性做一個探討,發現其是一個非常穩定的螯合劑,不會
因為酸鹼度(pH值)的變化或放置時間長短而改變其特性。接著探討8-HQ最
佳化的萃取濃縮條件,從研究中得知水樣的酸鹼度(pH值)、水樣流經8-HQ
的流速及沖滌液中鹽酸HCl和硝酸HNO3混合比例及濃度,皆是影響8-HQ萃
取濃縮條件。尋找出各個條件的最佳狀況即可建立一個精準的分析方法來
檢測出河水中的重金屬含量。

在石墨爐式原子吸收光譜儀中,在Zeeman背景校正系統下,尋找最佳的昇
溫程式來直測Cu、Cd的含量。可從改變四個升溫步驟(乾燥、灰化、原子
化、清除)的溫度,及在水樣中添加修飾劑如Mg(NO3)2或Pd-Mg來增加灰化
步驟的溫度,以其能完全去除水樣中的干擾物質,來達到重金屬在原子化
的步驟時可完全被偵測到,則原子吸收光譜儀有最佳的偵測條件。

本文研究發展的分析方法,應用8-HQ和GAAS,不僅能適用於河口水的重金
屬分析,對於海水中之重金屬含量,亦能適用之。所以,本法可以廣泛應
用於水污染防治,評估河口及沿岸環境品質。



Abstract

In this study, 8-Hydroxyquinoline (8-HQ) strategy was utilized
to extract heavy metals in estuarine water . The concentrated
sample was then further analyzed for Cu, Cd by Graphite atomic
absorption spectrometry (GAAS). After characterized the
synthesized 8-HQ, it found that 8-HQ chelating agent possesses
high stability against the fluctuation of the pH value and it
will not lose its chelating ability while it was exposed to the
testing environment. The optimized working condition of 8-HQ for
extraction hey metals in estuarine water was determined by
factors including the flowing rate of water sample through 8-HQ
as well as the different ratio of HNO3 and HCl contracted acid.
When all these factors were determined and then combined, the
right procedure for analyzing trace metals in estuarine water
was found.

Calibrating GAAS experiment in the Zeeman back-ground, finding
the correct temperature program for detecting the amount of
heavy metals could be simply achieved by changing the
temperatures of the four steps of the heating processes(drying,
ashing, atomization, cleaning, respectively). Furthermore,
adding modifier such as Mg(NO3)2 or Pd-Mg could increase the
temperature of ashing step leading to the elimination of any
undesired materials in the sample. With these modified steps,
all metals could be detectedhile their atomization process takes
place.

The analyzing method combined GAAS and 8-HQ is not only on the
study of trace metals in estuarine water but also on the
determination in sea water. So the method provide great
potential in coastal estuarine water pollution prevention
especially in detecting the existence of heavy metals which is
harmful to human.



QRCODE
 
 
 
 
 
                                                                                                                                                                                                                                                                                                                                                                                                               
第一頁 上一頁 下一頁 最後一頁 top